Das Dokument gilt für folgende Produkte:
1 Stück.
PDF-Vorschau wird geladen ...
As you inhale, the green feedback button on top of the nebulizer will move into the fully down position, indicating that the AeroEclipse* II BAN is producing aerosol in response to inhalation.
The AeroEclipse* II BAN can be operated in a Continuous Mode by rotating the mode selector 90 degrees clockwise to the continuous mode position ( ). The nebulizer can be returned to the Breath Actuated Mode by rotating the mode selector 90 degrees counterclockwise to the breath actuated position ( ). Cleaning Instructions Rotate the mode selector to the breath actuated position (green feedback button is up, mode selector has rotated counterclockwise) for both the washing and drying cycles.
CEN 13544-1 Data The Aerosol Output Rate for the AeroEclipse* II BAN is 0.16 ± 0.01 mL/min and the Aerosol Output is 0.51 ± 0.03 mL in Breath Actuated Mode.
Indications for Use The AeroEclipse* II BAN is intended to be used by patients who are under the care or treatment of a licensed healthcare provider.
• The AeroEclipse* II Breath Actuated Nebulizer can deliver aerosolized medication in either a Breath Actuated or Continuous Mode.
com * Continuous Mode Mode Selector Nebulizer Top Feedback Button C um m ua (...);
Datum der Aufnahme oder AktualiSierung des Dokuments: vor ca. 6 Jahren.
Qülle: Produzent Webseite [link].